1) gas flow error
[电]气流误差
2) electric error
电气误差
1.
In order to find the major error factors which affect the instrument accuracy, the reasons causing electric error are analyzed, the electric error consist of zero voltage error and power error.
为了确定影响仪器测量精度的主要误差因素,本文讨论了时栅传感器电气误差中零电平误差、电源误差的产生原因,提出用误差补偿技术来提高时栅的测量精度,对时栅传感器的批量化生产具有重要作用。
3) current error
电流误差
1.
The low current error and measures were discussed also.
还讨论了小电流误差的问题及对策。
2.
Compared with traditional methods,a value proportional to source voltage is added to the reference voltage,which counteracts the influence of the source voltage, improves the precision of APF output voltage, makes the output current follow the reference current more precisely,and reduce the range of output current error,in favor of reducing the power .
与传统方法相比,通过在输出指令电压中增加一个与电源电压瞬时值成正比的量,抵消了电源电压对输出指令电压的干扰作用,提高了APF输出电压的准确性,使APF的实际输出电流能够更加精确地跟踪指令电流的变化,从而有效地减小了输出电流误差的脉动范围,这也有利于减小有源电力滤波器的工作损耗和整机容量。
4) low current error
小电流误差
5) motor current error
电机电流误差
6) zero-error current tracking
零误差电流跟踪
1.
Based on the principle of internal model control, the grid-connected control with zero-error current tracking is appl.
基于内模控制原理,逆变器采用零误差电流跟踪并网控制方法,与普通PI控制方法相比,采用该方法后,工频条件下输出电流可以零误差地跟踪给定电流,基本不受电网电压的影响。
补充资料:厂用电受电
厂用电受电
energizing of auxiliary power system
chongyongd一on sho一Jd一on厂用电受电(energizing of auxiliary powersystem)机组附属设备安装完毕后需通电试运行,所需厂用电源由外部供给,因此厂用电受电是调试工作开始的标志。 受电步骤新建电厂先由电力系统向变电站供电,再依次对高压起动/备用变压器、3~10kv厂用配电装置、厂用工作变压器及380V配电装置供电。扩建电厂因高压起动/备用变压器已经投人运行,故可利用它的电源对扩建的厂用配电装置等供电。 调试项目主要有:①用额定电压依次对空载线路、变电站母线、隔离开关、断瘩器、互感器、变压器及厂用配电装置等进行冲击合闸试验。有条件时.冲击合闸前应先进行递升加压试验。升压过程中注意检查各设备有无放电声及短路现象,发现问题及时处理后再进行升压。②检查三相电压应平衡,相序应正确,各段母线的相位彼此应一致。对配电装置供电时,有时会由于母线电容和与母线连接的电磁式电压互感器的电感相互作用而出现铁磁谐振,此时相电压升高,中性点明显位移。为消除谐振可在电压互感器的剩余电压绕组处并联电阻以去振。